Association rate constant (ka)Proteins 103to 3 x 109M-1s-1LMW molecules 103to 5 x 107M-1s-1
Automation48h unattended operation
Baseline drift0.3 RU/min
Baseline noise0.03 RU (RMS)
Buffer selectionAutomatic switching between four buffers
Capacity1 x 96- or 384-well microplate and up to 33 reagent vials or 78 vials for samples and reagents
Concentration rangeSample concentration → 1 pM
Dissociation rate constant (kd)10-5to 1 s-1
Flow cell volume60 nL
Flow rate1 to 100 μL/min
Injection volume2 to 350 μL
Instrument dimensions, W x H x D600 x 690 x 615 mm
Mains requirementProcessing unit: Autorange 100 to 240 V~, 50/60 Hz
Molecular weight detectionNo lower limit for organic molecules
Number of flow cells4
Power consumptionProcessing unit: max. 400 VA
Refractive index1.33 to 1.40
Sample storage temperature4°C to 45°C (maximum 15°C below ambient temperature)
Sample typeSmall-molecule drug candidates to high-molecular weight proteins (also DNA, RNA, polysaccharides, lipids, cells, and viruses) in various sample environments (e.g., in DMSO-containing buffers, plasma, and serum)
Sample volume min.Sample volume Injection volume plus 20 to 50 µL (application dependent)