Model | - Mettler Toledo, 2018년 5월 - Compact KF Coulometers, C20S - Gas phase extraction oven, DO308 |
---|---|
Operating time | 09:00 ~ 18:00 |
Location | IBS HQ C179 Chemical Process lab |
inquiry |
1. Equipment name : Karl-Fisher titrator (칼피셔 수분측정기)
2. Manufacture and Model
- Mettler Toledo, 2018년 5월
- Compact KF Coulometers, C20S
- Gas phase extraction oven, DO308
3. Purpose
- To analyze of water concent in a sample from any industry with water content between 1 ppm and 5%
4. Specification and performance
1. Compact KF Coulometers, C20S
- Polarized sensor input
Measurement range : - 2000 mV ... + 2000 mV
Resolution / Limits of error : 0.1 mV / 2 mV
Polarization current range : 0 ... 24 uA
Polarization current resolution / Limits of error: 0.1 uA / 1 uA
- Environment
Ambient temperature : + 5 ... + 40 °C
Atmospheric humidity : max. 80% at 31 °C
- Titration dimensions
WxDxH / weight: 210×340×312 mm / 3.3 kg
Glass beaker : 200 mL
- Control and display
Touchscreen : VGA 5.7" color TFT, 640×480 pixel
Application
Drift measurement : Online, < 2 ug/min
Measurement range : 10 ug ... 200 mg(1 ppm ~ 5%)
Resolution : 0.1 ug
Repeatability : 0.3% at > 1 mg H2O
2. Gas phase extraction oven, DO308
- Temperature range : 40 ~ 300 °C
- Temperature accuracy : ± 5 °C
- Dimensions overall: 300 (W) x 300 (D) x 550 (H) mm
- Weight : 9.1 kg
5. Location and Picture
- IBS headquarter, C179 Chemical Process lab