To explore new physics phenomena of low dimensional materials
with a special emphasis on two-dimensional layered structures
Model | CRX-VF 외 5 |
---|---|
Operating time |
SUN(00:00~24:00) MON(00:00~24:00) TUE(00:00~24:00) WED(00:00~24:00) THU(00:00~24:00) |
Location | 86691 |
inquiry |
Nguyen Tuan Dung 010-9486-2804 |
Available Time // Except - Every weeks Fri. 9:00~24:00, Sat 0:00~24:00.
Notice // If you want to have training for using this equipment, please contact Super-user.
Current Status // CCR is OK. You can use B-field and IV.(~2016/02/05)
Measurement of electrical properties of materials.
It is equipped with a vertical field superconductiong magnet capable of a maximum 2.25T.
It provides various temperature operation (10K to 500K) without liquid He.
Magnetic Field : Base (<10K) :-2.5T~+2.5T , 10K to 400K : -2T~+2T , 400K to 500K : -1T ~ +1T .
Cool down time(Room temperature to within 10K of base) : 4.5h .
Warm up time (Entire system from base to room terperature) : 2h. Pump Down Time : Before CCR start-up(10^-3Torr) 1.5h .
Vacuum : Room temperaute <5x10^-4 Torr , at base temperature <1x10^-5 Torr , at maximum temperature <5x10^-5 Torr.
Sample area : Up to 51mm (2 in) .
Sample Thinkness: Up to 12mm. Working height of sample: 897mm . Tip materials: Tungsten( radius 10um, 25um) , BeCu (radius 10um) .